V
Scaner-VS
HomeCatalogSourcesCWECAPECATT&CKMitigationsDocs
CVE-2018-10596
CVE
High

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still c…

CVSS
7.1
High
EPSS
0.00
p39
Published
2018-01-01
Updated
2018-01-01
Description

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

Tags · CWE
CWE-923
CAPEC-161
CAPEC-481
CAPEC-501
CAPEC-697
Affected products
2090_carelink_programmer_firmware
CVSS vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
Timeline
2018-01-01
Published
2018-01-01
Updated
CVSS 3.1 breakdown
Attack Vector
AV: N
Network (N)
Attack Complexity
AC: H
High (H)
Privileges Required
PR: L
Low (L)
User Interaction
UI: R
Required (R)
Scope
S: U
Unchanged (U)
Confidentiality Impact
C: H
High (H)
Integrity Impact
I: H
High (H)
Availability Impact
A: H
High (H)
Exploit indicators
EPSS
0.002 · p39
Known exploited (KEV)
No
MITRE ATT&CK
Inferred via CAPEC
└ via CAPEC-481 · CWE-923
└ via CAPEC-697 · CWE-923
Known exploits — Сканер-ВС
No Сканер-ВС checks registered for this vulnerability yet.
Affected software
ProductVendorStatus
2090_carelink_programmer_firmware*Tracked
Source databases
CVE